Number

4

Design

galvanically isolated, single ended

can be configured: voltage (10 V) or current (20 mA)

Input signal range

10 V

-10 V ...+10 V

20 mA

-20 mA …+20 mA

Max. input voltage

±60 V permanent

Resolution

16 bit

Sampling rate

max. 40 kHz, freely adjustable1

Filter

permanent

R/C low-pass, 1st order, 40 kHz

in addition

analog anti-aliasing filter (Butterworth), 4th order, 20 kHz

digital anti-aliasing filter (Tschebyscheff I)2, 8th order, cutoff frequency can be adjusted

Input resistance

voltage (U)

device switched off

80.6 kΩ

device switched on

107.2 kΩ

current (I)

device switched off

50 kΩ

device switched on

50 kΩ

Input capacity

voltage (U)

4.5 pF

current (I)

n/a

Frequency range

0 Hz… 20 kHz

Accuracy

< 0.1 % of total measuring range

Electrical isolation

Channel-channel

1.5 kV AC

Channel - housing/power supply

1.5 kV AC

Connector type

1x 12-pin multi-pin connector; screw-type terminal (0.14 mm2 to 1.5 mm2), screw connection, included in delivery

Additional functions

Power frequency measurement

10 Hz… 80 Hz

interval

1 s / 10 s (according to DIN EN 61000-4-30)