The measuring ranges are only valid for the ibaMS3xAI-1A/100A module. The internal 18 bit processing of the ibaMS3xAI module allows to display different measuring ranges in high resolution with 16 bit.
|
Measuring range |
Resolution |
Calibrated |
Analog filter |
Description |
|---|---|---|---|---|
|
1.56 A |
16 bit |
x |
x |
for currents < 1.56 A DC |
|
6.25 A |
16 bit |
x |
x |
default |
|
100 A |
16 bit |
x |
- |
for currents < 100 A DC (only for 1 second) |
|
6.25 A / 100 A |
15 bit |
x /x |
x / - |
combined range with 6.25 A and 100 A range. Switch-over at 6.20 A. |
Note |
|
|---|---|
|
The additional analog filters can be adjusted only for the 6.25 A and the 1.56 A ranges! |
|