Caution! |
|
|---|---|
|
You must only connect one conductor to each terminal connection. Several individual conductors, whether single-wired or fine-wired, are not permitted. Only connectors classified by iba may be used for connecting conductors. |
|
|
Connection technology |
Push-in |
||||
|
Clamping range |
0.13 - 3.31 mm2 |
||||
|
Conductor cross-sections |
|||||
|
Single-wired |
0.2 - 2.5 mm2 |
||||
|
Fine-wired |
0.2 - 2.5 mm2 |
||||
|
With wire end ferrule |
0.25 - 2.5 mm2 |
||||
|
With wire end ferrule/collar |
0.25 - 2.5 mm2 |
||||
|
Stripping length |
|||||
|
Cross-section |
0.5 mm2 |
0.75 mm2 |
1 mm2 |
1.5 mm2 |
2.5 mm2 |
|
Single-wired |
10 mm |
||||
|
Fine-wired |
|||||
|
With wire end ferrule |
10 mm |
||||
|
With wire end ferrule/collar |
12 mm |
n/a |
|||
|
Recommended cables |
|||||
|
Single-wired |
H05V-U; H07V-U |
||||
|
Fine-wired |
H05V-K; H07V-K |
||||
|
Screwdriver blade |
0.6 mm x 3.5 mm |
||||
Caution! |
|
|---|---|
|
Measuring cable
|
|
Note on the maximum sensor cable length for IEPE modes
The maximum length of the sensor cable depends on various factors, including the set sampling rate. The following formula can be used to estimate the maximum length.
|
|
Sampling rate [Hz] |
|
|
Capacity [F/m] |
|
|
Maximum output voltage of the sensor [V] |
|
|
Power supply for IEPE evaluation electronics [A] |
Sample calculation
|
Cable LAPP Unitronic FD CP (TP) plus 1x2x0.75mm2 |
|
|
Sensor IMI Piezotronics Type 601A01 |
|
|
Evaluation electronics iba AG (IEPE module) |
|
|
Set sampling rate |
|
Result: l
max = 45,5 m