Danger! |
|
|---|---|
|
Operation
|
|
Danger! |
|
|---|---|
|
Permanent measuring voltage The device is non-destructive for permanent measuring voltages up to 60 V DC. The device can be destroyed at voltages from 60 V DC and device safety can no longer be guaranteed. |
|
Danger! |
|
|---|---|
|
Duty of care Take care when working on the system and always check that the system and the modules themselves are in perfect condition, as well as ensuring that they are properly installed and correctly attached to the DIN rail. If damage to cables, devices, supplies or enclosures is detected before commissioning or during operation, the system must not be put into operation or must be taken out of operation immediately. |
|
Warning! |
|
|---|---|
|
Mounting and dismounting / Disconnection from the grid Work on the device or system may only be carried out when the power is switched off! Due to the modular concept of this system, modules connected in series with this module can also carry dangerous voltages. All energized components of all modules in the system must therefore be disconnected from the grid before mounting and dismounting. In addition to disconnecting the power supply at the system's central unit, the signal plugs and connections of all modules in the system must also be de-energized or disconnected from the grid. |
|
Caution! |
|
|---|---|
|
A suitable disconnecting device for this system must be available and disconnect all energized components of this system. This disconnecting device must include a switch or circuit breaker that is easily accessible at a suitable location in the vicinity and is also clearly marked as a disconnecting device for this system. |
|
Caution! |
|
|---|---|
|
Measuring cable
|
|
For information on the maximum length of the measuring cable, see chapter Connector connection technology.
Caution! |
|
|---|---|
|
You must only connect one conductor to each terminal connection. Several individual conductors, whether single-wired or fine-wired, are not permitted. Only connectors classified by iba may be used for connecting conductors. |
|
Caution! |
|
|---|---|
|
If the display of an analog input lights up red, the input signal is outside the displayable and permissible range.
The actual voltage at the input is higher than the permitted voltage. |
|
Note |
|
|---|---|
|
Do not open the device! Opening the device results in a loss of warranty! |
|
Note |
|
|---|---|
|
The device does not require any special cleaning or maintenance! However, if you want to carry out an inspection or recalibration, return the device to iba. |
|
Note |
|
|---|---|
|
Calibration If the device is used or operated in the manner specified in the Technical data chapter, a calibration interval of 4 years is recommended for the analog input channels. The date of the last calibration can be found in ibaPDA and there in the system information on the Info tab of the central processing unit. |
|