The device is an electrical equipment. It must only be used for the following applications:
-
Measurement data acquisition and analysis
-
Applications of software products (ibaPDA, ibaLogic etc.) and hardware products from iba AG.
The device must only be used as specified in the Technical data chapter, and is designed and approved for continuous operation.
Danger! |
|
|---|---|
|
Electric shock The device is only designed for electrical measured variables as specified in the “Technical data” chapter! If the device is used or operated in a manner other than specified in the Technical data chapter, the protection supported by the device may be impaired. This applies in particular to the permissible operating and environmental conditions and voltages outside the corresponding CAT protection classes. |
|
Caution! |
|
|---|---|
|
This module is specially designed and developed for the use or input-side connection of protection transformers and for an input current/nominal current of 5 A AC. Even if this module can measure higher currents for short periods (150 A AC for 1 s per minute), it is still only designed for a continuous current load of 15 A DC. The use of protection transformers always involves a conceptual switch-off or protection of the measuring circuit in the event of an overcurrent on the protection transformer side. If currents are measured by standard transducers or current transformers or directly at the inputs, this module may only be operated if the measuring circuit is fused up to a continuous current of 15 A AC. |
|