Analog inputs

Number

4

Design

galvanically isolated, single ended

Sampling method

parallel

Input signal / nominal voltage

150 V AC

Resolution

24 bit (Delta-Sigma)

ADC sampling rate

500 kHz

ibaPDA timebase (update time)

min. 2 μs; freely adjustable (integer multiple); max. 1 ms for filter signals

Conversion time

not relevant; excluded during data acquisition

Filter

    ADC signal

Analog

R/C low-pass, 1st order, typ. 150 kHz

Digital

anti-aliasing filter (FIR), 84th order; cutoff frequency = 0.45 x ADC sampling rate; oversampling = 32 x ADC sampling rate

    Filter-Signal1

    Like ADC signal, in addition

digital

anti-aliasing filter (Elliptic/Cauer); 10th order; cutoff frequency = 0.45 / timebase

Measuring range

±312 V

Measuring category

CAT III 150 V

Input impedance

Device switched off

250 kOhm

Device switched on

251 kOhm

Input capacity

8.5 pF

Accuracy (+25 °C)

< 0.1 % of the respective double full scale value

Electrical isolation

Channel - channel

basic insulation: CAT III 150V

Channel - system

enhanced insulation: CAT III 150V

Connection technology

2x 4-pin header, pitch 7.62 mm

Connectors

2x enclosed; push-in, conductor max. 2.5 mm2, locking lever (latching), protected against reverse polarity, lockable; for information on the conductor and stripping length, see chapter Connection technology connector; order number: 52.000050

Additional functions

Phasor measurement unit2

integrated

Power frequency measurement (10 Hz ... 80 Hz)

interval 1 s / 10 s (according to IEC 61000-4-30)